24

Tracer AV Installation Manual

6. With

WVAL

displayed, press

Zero

to show

t h e st o r e d c a l i b r a t i o n w e i g h t . U s e t h e
procedure shown in Figure 3-3 12 to enter the
actual value of the test weights to be used for
the span calibration. Press

Zero

again

to

store the entered

WVAL

value and go to span

calibration (

WSPAN

).

7. With

WSPAN

displayed, place test weights on

the scale and press

Zero

to calibrate span.

The indicator displays

*CAL*

while calibration

is in progress. When complete, the A/D count
for the span calibration is displayed. Press

Zero

again to store the span calibration

value.

4.2

Standard Calibration

To calibrate the

Tracer AV

using the

Zero

and

Units

keys, do the following:

1. Place the

Tracer AV

in setup mode

To enter setup mode, remove the setup switch
access screw on the

Tracer AV

(See Figure 2-3

4). Once the screw is removed, use a small
screwdriver or pen tip to press the setup
switch. The display changes to show the word

DEFCAL

.

2. Remove all weight from the scale platform.

If your test weights require hooks or other
items, place them on the scale for zero
calibration.

3. Press the

Units

key until the display reads

CALIBR

(see Figure 4-1). Press the

Zero

key

to go to zero calibration (

WZERO

).

4. With

WZERO

displayed, press

Zero

to

calibrate zero. The indicator displays

*CAL*

while calibration is in progress. When
c o m p l e t e , t h e A / D c o u n t f o r t h e z e r o
calibration is displayed. DO NOT adjust this
value after WSPAN has been set!
Press

Zero

again to save the zero calibration value and go
to the next prompt (

WVAL

).

5. With

WVAL

displayed, press

Zero

to show

t h e st o r e d c a l i b r a t i o n w e i g h t . U s e t h e
procedure shown in Figure 3-3 12 to enter the
actual value of the test weights to be used for
the span calibration. Press

Zero

again

to

store the entered

WVAL

value and go to span

calibration (

WSPAN

).

6. With

WSPAN

displayed, place test weights on

the scale and press

Zero

to calibrate span.

The indicator displays

*CAL*

while calibration

is in progress. When complete, the A/D count
for the span calibration is displayed. Press

Zero

again to store the span calibration

value and go to the next prompt (

REZERO

).

7. The rezero function is used to remove a

calibration offset when hooks or other items
are used to hang the test weights.

If no other apparatus was used to hang the
test weights during calibration, remove
the test weights and press

Units

to

return to the CALIBR menu.

If hooks or other items were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, press

Zero

to rezero the scale.

This function adjusts the zero and span
calibration values. The indicator displays

*CAL*

while the zero and span calibrations

are adjusted. When complete, the adjusted
A/D count for the zero calibration is
displayed. Press

Zero

to save the value,

t h e n p r e s s

U n i t s

t o r e t u r n t o t h e

CALIBR menu.

8. Press

Units

until the display reads

CONFIG

,

and the display will exit the setup mode.

4.3

EDP Command Calibration

To calibrate the

Tracer AV

using EDP commands, the

EDP port must be connected to a terminal or personal
computer. See Section 2.3.3 on page 7 for EDP port
pin assignments; see Section 5.0 on page 26 for more
information about using EDP commands.

Once the

Tracer AV

is connected to the sending device,

do the following:

1. Place the

Tracer AV

in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require
hooks or other items, place them on the scale
for zero calibration.

2. Send the WZERO EDP command to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress.

3. Place test weights on the scale and use the

WVAL command to enter the test weight
value in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress.

5. To remove an offset value, clear all weight

from the scale, including hooks or other items
used to hang test weights, then send the
REZERO EDP command. The indicator
displays

* C A L *

while the zero and span

calibrations are adjusted.

6. Send the KUPARROW EDP command to exit

setup mode.