Tracer AV Installation Manual - Calibration

25

4.4

Revolution

®

Calibration

To calibrate the

Tracer AV

using

Revolution

, the EDP

p o r t m u s t b e c o n n e c t e d t o a P C r u n n i n g t h e

Revolution

configuration utility.

Use the following procedure to calibrate the

Tracer

AV

:

1. Select

Calibration Wizard

from the

Revolution

tools menu.

2.

Revolution

uploads calibration data from the

Tracer AV

then presents the information in a

display like that shown in Figure 4-3.

Figure 4-3.

Revolution Calibration Display

3. Enter the

Value of Test Weight

to be used for

span calibration then click

OK

.

4. The Zero Calibration dialog box prompts you

to remove all weight from the scale. Clear the
scale and click

OK

to begin zero calibration.

NOTE: If your test weights require hooks or other items,
place them on the scale for zero calibration.

5. When zero calibration is complete, the Span

Calibration dialog box prompts you to place
test weights on the scale for span calibration.
Place tests weights on the scale then click

OK

.

6. When calibration is complete, the

New Settings

fields of the Indicator Calibration display are
filled in. Click

Exit

to save the new values and

return to the

Revolution

main menu; to restore

the previous calibration values, click

Restore

Settings

.

4.5

More About Calibration

The following topics provide additional information
about compensating for environmental factors
(Section 4.5.1) and diagnostic information for
determining expected zero and span coefficients.

4.5.1

Adjusting Final Calibration

Calibration may be affected by environmental factors
including wind, vibration, and angular loading. For
example, if the scale is calibrated with 1000 lb, a
s t r a i n t e s t m a y d e t e r m i n e t h a t a t 2 0 0 0 l b t h e
calibration is 3 lb high. In this case, final calibration
can be adjusted by tweaking WVAL to 998.5 lb. This
adjustment provides a linear correction of 1.5 lb per
1000 lb.

To adjust the final calibration, return to the

WVAL

prompt and press Zero to show the test weight
value. Press

Units

to adjust calibration up or down.

Press Zero to save the value, then press

Units

to

return to the CALIBR menu.

4.5.2

A/D Counts

Table 4-3 lists the ideal A/D counts that result from
input signals of 0–3.0 mV with zero deadload. Actual
values will typically be higher than the values shown
in Table 4-3, but the ideal values can be used when
calibrating the

Tracer AV

with no attached scale.

NOTE: When Raw A/D Count is displayed, the six most
significant digits appear. Scroll left or right to see the full
value.

Input Signal (mV)

Raw A/D Counts

0.0 mV/V

167,840

0.5 mV/V

335,613

1.0 mV/V

503,377

1.5 mV/V

671,143

2.0 mV/V

838,908

2.5 mV/V

1,100,668

3.0 mV/V

1,174,446

Table 4-3. Ideal A/D Raw Counts