Press the, Softkey – Rice Lake BCi Belt Scale - Calibration Guide User Manual

Page 12

Advertising
background image

8

BCi In-Motion Belt Scale System

3. Press the

Start

softkey

5/30/2013

09:18AM

Exit =>

Start

SCALE #1

Progress

Test Time

Test Accumulator

Old Span Error %

Span Error %

6.2

0.00

2.55

0.00%

Figure 3-14. Start Auto Cal with Test Weights

4. The Auto Cal test will run, and progress is shown on the screen. A prompt to enter a test weight value will

display.

5/30/2013

09:35AM

End

Home

SCALE #1

Progress

Test Time

Test Accumulator

Old Span Error %

Span Error %

18.8

62

0.00

Enter Test Weight Value
=> 10.0

Cancel

Figure 3-15. Enter Test Weight Value

5. Press the

CLR

key on the integrator to clear off the current value, enter a new test weight value using the

numeric keypad.

6. Press the

enter

key on the integrator to accept the next test weight value. After the test the test weight value

is entered, the new span error is displayed.

7. Press the

Yes

softkey to accept that value.

8. Repeat the test until the new span percent error is very near zero and then accept the new span value by

pressing the

Yes

softkey.

5/30/2013

09:32AM

No

Yes

SCALE #1

Progress

Test Time

Test Accumulator

Old Span Error %

Span Error %

18.8

62

95.00

0.10

Accept New Span Value

Figure 3-16. Accept Span Value

9. Press the

Exit

softkey to exit back to the integrator weigh screen.