BCi In-Motion Belt Scale System

3. Press the

Start

softkey. This will cause the integrator to change screens displaying the number of pulses

counted, the frequency of pulses and the time the test is running, Figure 3-4.

5/30/2013

08:16AM

Exit=>

Start

SCALE #1

Time Running

Finish

769 Pulses

122 Per Sec

6.2

Figure 3-4. Start Softkey

The screen illustrates a live display of how the belt is working. If the pulses and frequency are not
changing then there is a problem with the speed sensor and can be a sign that either the wiring or the sensor
itself is bad. During this cycle of the calibration procedure, the operator needs to count the number of belt
revolutions.

4. Press the

Finish

softkey when the marks are properly aligned and this will stop the speed sensor calibration.

The integrator will then prompt the user to

Enter The Number of Revolutions

(Figure 3-5).

5. The default number is 3 and to change this number, press the

CLR

key to erase that number and then enter a

numeric value from the integrator keypad to enter the new number of revolutions.

5/30/2013

08:16AM

End

Home

SCALE #1

Time Running

Cancel

2282 Pulses

122 Per Sec

18.6

Enter Number of Revolutions
=> 3

Figure 3-5. Enter Number of Revolutions

6. Press the

enter

key on the integrator to save the value and return to the main supervisor menu.