Rice Lake BCi Belt Scale - Calibration Guide User Manual
Page 8
4
BCi In-Motion Belt Scale System
3. Press the
Start
softkey. This will cause the integrator to change screens displaying the number of pulses
counted, the frequency of pulses and the time the test is running, Figure 3-4.
5/30/2013
08:16AM
Exit=>
Start
SCALE #1
Time Running
Finish
769 Pulses
122 Per Sec
6.2
Figure 3-4. Start Softkey
The screen illustrates a live display of how the belt is working. If the pulses and frequency are not
changing then there is a problem with the speed sensor and can be a sign that either the wiring or the sensor
itself is bad. During this cycle of the calibration procedure, the operator needs to count the number of belt
revolutions.
4. Press the
Finish
softkey when the marks are properly aligned and this will stop the speed sensor calibration.
The integrator will then prompt the user to
Enter The Number of Revolutions
5. The default number is 3 and to change this number, press the
CLR
key to erase that number and then enter a
numeric value from the integrator keypad to enter the new number of revolutions.
5/30/2013
08:16AM
End
Home
SCALE #1
Time Running
Cancel
2282 Pulses
122 Per Sec
18.6
Enter Number of Revolutions
=> 3
Figure 3-5. Enter Number of Revolutions
6. Press the
enter
key on the integrator to save the value and return to the main supervisor menu.