2 integrator calibration, Zero cal, 1 zero cal – Rice Lake BCi Belt Scale - Calibration Guide User Manual
Page 9: Softkey, Press the, Softkey on the integrator
Calibration
5
3.2
Integrator Calibration
There are three modes of integrator calibration.
• Zero Cal
• Auto Cal
• Material Cal
All three calibration methods are described in the following sections.
3.2.1
Zero Cal
By selecting this mode it will calibrate the integrator based on a zero test. A zero calibration is based on the number
of belt revolutions is established during the belt speed sensor calibration.
Use the following steps to perform a Zero Cal calibration.
1. To access the mentioned softkeys, press the
Integrator Calibration
softkey.