2 integrator calibration, Zero cal, 1 zero cal – Rice Lake BCi Belt Scale - Calibration Guide User Manual

Page 9: Softkey, Press the, Softkey on the integrator

Advertising
background image

Calibration

5

3.2

Integrator Calibration

There are three modes of integrator calibration.

• Zero Cal

• Auto Cal

• Material Cal

All three calibration methods are described in the following sections.

3.2.1

Zero Cal

By selecting this mode it will calibrate the integrator based on a zero test. A zero calibration is based on the number
of belt revolutions is established during the belt speed sensor calibration.

Use the following steps to perform a Zero Cal calibration.

1. To access the mentioned softkeys, press the

Integrator Calibration

softkey.

Exit=>

SCALE #1

Master

Totalizer

Integrator

Calibrati

5/30/2013

08:13AM

Admin. Passcode (numeric)

Scale Capacity (Max Rate)

Load Cell MV

Total LC Build

Rate Unit Time

Filter (s)

Filter Threshold (divisions)

Speed Unit Time

Fixed Speed

Unit of Measure

420.00

1.99900

800

hr
7.0

2

min

0.00

ft

Belt

Calibration

Time/Date

Figure 3-6. Integrator Calibration Softkey

2. Press the

Zero Cal

softkey on the integrator.

Exit=>

SCALE #1

Auto Cal

5/30/2013

08:18AM

Admin. Passcode (numeric)

Scale Capacity (Max Rate)

Load Cell MV

Total LC Build

Rate Unit Time

Filter (s)

Filter Threshold (divisions)

Speed Unit Time

Fixed Speed

Unit of Measure

420.00

1.99900

800

hr
7.0

2

min

0.00

ft

Zero Cal

Material

Figure 3-7. Zero Cal Softkey