BCi In-Motion Belt Scale System

15. Press the

Yes

softkey to accept that value.

16. Repeat the test until the new span percent error is very near zero and then accept the new span value by

pressing the

Yes

softkey.

5/30/2013

09:32AM

No

Yes

SCALE #1

Progress

Test Time

Test Accumulator

Old Span Error %

Span Error %

18.8

62

95.00

0.10

Accept New Span Value

Figure 3-20. Accept New Span Value

17. Press

Exit

until back at the integrator weigh screen.

3.2.3

Material Calibration

Use this calibration mode to calibrate the scale with a known amount of material. The material must be
pre-weighed or post weighed.

To perform a material calibration, choose the

Material Cal

softkey and follow the same steps as a Span Calibration

and instead of entering a test weight value, enter the weight of the material used in tons.