Material calibration, 3 material calibration – Rice Lake BCi Belt Scale - Calibration Guide User Manual
Page 14
10
BCi In-Motion Belt Scale System
15. Press the
Yes
softkey to accept that value.
16. Repeat the test until the new span percent error is very near zero and then accept the new span value by
pressing the
Yes
softkey.
5/30/2013
09:32AM
No
Yes
SCALE #1
Progress
Test Time
Test Accumulator
Old Span Error %
Span Error %
18.8
62
95.00
0.10
Accept New Span Value
Figure 3-20. Accept New Span Value
17. Press
Exit
until back at the integrator weigh screen.
3.2.3
Material Calibration
Use this calibration mode to calibrate the scale with a known amount of material. The material must be
pre-weighed or post weighed.
To perform a material calibration, choose the
Material Cal
softkey and follow the same steps as a Span Calibration
and instead of entering a test weight value, enter the weight of the material used in tons.