2 weighing module, 1 types, 2 technical specifications – Rice Lake Contego-Hygienic Weigh Module Kits, Lateral Cable Outlet User Manual
Page 10: Weighing module, Types, Technical specifications
Hygienic Weighing Module Contego®
Installation Manual
EN-8
Sartorius
3.2
Weighing Module
3.2.1
Types
3.2.1.1
Weighing Module Incl. Load Cell
For nominal load E
max
Error class
Type*
100 kg
D1
A, Ex A, B, Ex B
200 kg
D1, C3
A, Ex A, B, Ex B
300 kg
D1, C3
A, Ex A, B, Ex B
500 kg
D1, C3
A, Ex A, B, Ex B
1 t
D1, C3
A, Ex A, B, Ex B
2 t
D1, C3
A, Ex A, B, Ex B
* A = cable outlet on side
B = cable outlet on top
Ex = for installation in the explosion-hazarded area
3.2.1.2
Weighing Module without Load Cell (Service)
PR 6041/60S
Mounting kit with cable outlet on side
PR 6041/70S
Mounting kit with cable outlet on top
3.2.2
Technical Specifications
Permissible horizontal force
7 kN
Permissible vertical load without load cell
1.5 t
Permissible lift-off force
8 kN
Permissible force for jack-up
15 kN
Permissible temperature range
-40 °C to +95 °C
Protection classes
Dust and water
IP69K:
Protection against dust
as per DIN EN 60529 09/2000 Paragraph 13.4
Protection against water (high-pressure cleaning)
as per DIN 40050 Part 9
Explosion
Suitable for sub-groups IIA, IIB, and III for explosive
atmospheres. Please ask regarding IIC compliance.
Material
Bellow
Silicone (FDA-compliant)
Mounting kit
1.4404 (AISI 316L)
Weight of
cpl. weighing module
net
approx. 11 kg
gross
approx. 12 kg