Installation Manual

Hygienic Weighing Module Contego®

Sartorius

EN-17


Note:

The following operations must be performed at all supporting points (e.g. vessel foot) of
the weighing object (e.g. vessel)!

Procedure

Caution:
The following contact surfaces must be clean and even to ensure reliable sealing:
-

between upper plate and vessel foot

-

between lower plate, adaptor plate (if any) and foundation

• Check, if all threaded holes required for mounting (see Chapter 4.1.2) in the vessel foot (6) are

provided.

• Insert the O-ring (5) into the groove of the upper plate (4) and put the weighing module under the

vessel foot (6).

• Slip the bellow (not shown) over the mounting kit and the vessel foot (6) carefully so that the

mounting kit remains accessible for mounting.

• Fit the mounting kit to the vessel foot (6). It is essential to observe the property classes and tightening

torques of bolts (2) and washers (1); refer to Chapter 4.1.2.