Hygienic Weighing Module Contego®

Installation Manual

EN-20

Sartorius

5

Connection

5.1

General Information

Caution:
The load cell cable end is protected against contamination and moisture by a
protective sheathing.

Remove the protective sheathing only shortly before connecting!

• Protect the cable ends against contamination. No moisture must penetrate into the open cable end.
• Don't shorten the load cell connecting cable. Connect the prepared cable end and roll up the remaining

cable.

• The cable screen may be connected only to the connecting terminals of the indicator.
• Keep the load cell cables away from power cables.
• The distance between the measuring cables and the power cables or components under high voltage

should be at least 1 m (reference value).

• We recommend laying the load cell cables in separate cable trays or armored steel pipes.
• Power cables should be crossed at right angles.

5.2

Load Cell

+ Supply voltage


+ Meas. voltage / + Load cell output

– Supply voltage

– Meas. voltage / – Load cell output

Screen

Colour Code

rd = red
bu = blue
gn = green
gy = grey