2 checks with multimeter, 1 check zero output signal of load cell, 4 check insulation impedance of connecting cable – Rice Lake Contego-Hygienic Weigh Module Kits, Lateral Cable Outlet User Manual

Page 27: Checks with multimeter, Check zero output signal of load cell, Checking the strain gauge bridge of the load cell, Checking the insulation impedance of the load cell, Check insulation impedance of connecting cable

Advertising
background image

Installation Manual

Hygienic Weighing Module Contego®

Sartorius

EN-25

7.2

Checks with Multimeter

7.2.1

Check Zero Output Signal of Load Cell

• Unload load cell.
• Disconnect the load cell measuring outputs.
• Check whether the output voltage without load is within the limits.

Type

Zero output signal

D1, C3

0 mV ±0.02 mV/V

7.2.2

Checking the Strain Gauge Bridge of the Load Cell

• Do not exceed the test voltage.
• Check whether the values of the resistors are within the permissible limits.

Maximum test voltage

Standard industrial

28 V DC

intrinsically safe version (PR …/…E)

25 V DC

Type

Input impedance
(red core, blue core)

Output impedance
(green core, gray core)

D1

650

Ω ±6 Ω

610

Ω ±1 Ω

C3

650

Ω ±6 Ω

610

Ω ±0.5 Ω

7.2.3

Checking the Insulation Impedance of the Load Cell

• Never apply the test voltage between the cores of the load cell cable (danger of destroying the load

cell).

• Insulate the load cell cores.

Maximum test voltage

Standard industrial

100 V DC

Intrinsically safe version (PR …/…E)

500 V AC

Insulation impedance

Core - housing

>5000 M

Core - screen

>5000 M

Screen - housing

>5000 M

7.2.4

Check Insulation Impedance of Connecting Cable

• Disconnect connecting cable and load cells.
• Insulate the cores of the connecting cable.

Insulation impedance

Core - core

≥600 MΩ × km

Core - screen

≥600 MΩ × km