Page 13: Calibration, Normal calibration linear calibration, Unload, F1 unt, Pi n, P 1 lin, Load 0
Normal Calibration
9
4.0
Calibration
4.1
Normal Calibration
1. Press and hold
to enter calibration.
unload
is display.
2. Remove any weight from the platform.
3. After the stable indicator (
) is displayed, press
to confirm. Load
will display.
4. Place the calibration weight (1/3 of the capacity) on the platform.
4.2
Linear Calibration
1. Turn on the balance. Press
during self test.
F1 UNT
will display.
2. Press
several times to scroll to tech, then press
to enter
calibration.
PI N
will display.
3. Press
.
P 1 LIN
will display.
4. Press
to enter calibration.
PI N
will display.
5. Press .
load 0
will display.
6. Ensure the platform is empty. After stable (
) and zero ( ) indicators
come on,
LOAD 1
will display.
7. Place first calibration weight (1/3 of the capacity) on the platform. After
stable (
) and zero ( ) indicators come on,
load 2
will display.
8. Place next calibration weight (2/3 of the capacity) on the platform. After
stable (
) and zero ( ) indicators come on,
load 3
will display.
9. Place the last calibration weight (full capacity) on the platform. After stable
(
) and zero ( ) indicators come on, balance will start self test and return
to normal weighing mode.
Zero
Zero
Mode
Mode
Mode
Mode
Zero
Sample
Sample
Tare
Zero
Sample
Sample