14 Equipoise Series Technical Manual

7.0

Maintenance

7.1

General

Balance does not operate properly

Determine whether the problem is constant or alternate. Be aware that problems can

be caused by mechanical or electrical influences.
Check the following:

• Water
• Corrosive materials
• Vibrations, temperature or wind
• Physical damage

Check the balance cables for damage; check all connections and connectors for any

loose contact or incorrect connection.

Cleaning

• Disconnect the power before cleaning.
• Use a clean and soft cloth with mild cleaning agents.
• Make sure that the fluid is not able get into the device.

7.2

Error Codes

Error

Code

Description

POSSIBLE CAUSES

Err 1

Date set error

Check input data

Err 2

Time set error

Check input data

Err 3

Calibrationweight is too light

Check calibration weight

Err 4

Zero range exceeded, due to turning on

or by pressing

Goods on the platform
Overload, when zeroing the
balance
Improper calibration
Load cell problem
PCB problem

Err 5

Keyboard error

Some key short or hold over 30
seconds

Err 6

A/D Count out of the range

Platform not installed
Load cell problem
PCB problem

Err 7

Pecentage setting error

0.01% weight must be >0.5d

Table 7-1. Error Codes