Rice Lake Equipoise Series, Rice Lake Weighing Systems User Manual
Page 19
Error Codes 15
Err 8
External calibration weight error
More than 4% error as linear
calibration
Err 9
Unstable, can’t perform zero or tare
Err 10
No data
RS-232, continous
communication
Err 11
Won’t accept communication protocal
Can get signal but can’t support
Err 12
Accumulation error
Max accmulate time 999 or total
weight 999999 has been
reached.
Err 13
Lack of unit weight
Counting: unit weight < 0.5d
Err 14
Lack of sample
Counting: sample weight < 20d
Err 15
Error gravity value setting
Gravity value range 0.96-1.04
Err 16
No paper
Err 17
Tare Operation
Minus weight or overload
Err 18
Incorrect pre-tare operation
Minus weight or overload
Err P
Printer no response 2 seconds later
OK
Set password OK
PASS
Calibration OK
Fail
Calibrate or password setting fail
NG
RS-232 Command error
--OL-
Overload
Max +9d (NTEP: 105% capacity)
--LO-
Minus weight
-5% capacity
Bat lo
Battery voltage too low
<5.9v
Measuring Points
Resistance
Red (+ Exc) to White ( –Exc)
420 ±20Ω
Green (+Sig) to Black ( –Sig)
350Ω ±5Ω
Error
Code
Description
POSSIBLE CAUSES
Table 7-1. Error Codes