CS750, Users

Rev R, May 2009

Page 14 of 27

Start up

1. Move the calibration blocking switch to the “CAL” position if you intend to

calibrate.

2. Turn scale power ON and wait for scale to warm up (3 minutes from power

on).

3. Press ZERO and hidden keys together and release to enter the calibration

mode.

First ten parameters

4. The scale shows “EE-EE

EE-EE

EE-EE

EE-EE”. Press the zero key. To skip to “EE-10

EE-10

EE-10

EE-10” enter a “1

11

1”.

To skip to “LL-00

LL-00

LL-00

LL-00” enter “2

22

2”. No skips will occur with an entry of “0

00

0”.

5. The scale shows “EE-00

EE-00

EE-00

EE-00”. Press the zero key. Enter the sample rate (1 to

64). The sample rate is the number of past readings that are averaged
together to make a reading.

6. The scale shows “EE-01

EE-01

EE-01

EE-01”. Press the zero key. Enter the update rate (1 to

32). The update rate is the speed at which the displayed weight is updated.
The smaller the number the faster the display will be updated.

7. The scale shows “EE-02

EE-02

EE-02

EE-02”. Press the zero key. Enter demand versus

continuous on the optional serial output. For this product, you should always
set this to "Continuous" (0).

Setting

Type

0

Continuous on

1

Continuous off

8. The scale shows “EE-03

EE-03

EE-03

EE-03”. Press the zero key. Enter the baud rate of the

serial output (0 to 7).

Setting

Baud Rate

0

9600

1

4800

2

2400

3

1200

4

600

5

300

6

150

7

75

8

19.2K

9

38.4K