Rice Lake Intercomp CS 750 - Users Manual User Manual

Page 17

Advertising
background image

CS750, Users

Rev R, May 2009

Page 17 of 27

18. The scale shows “EE-13

EE-13

EE-13

EE-13”. Press the zero key. The scale shows the current

initial zero range setting. Use the following table to select the “initial zero
range” setting:

Setting

Initial zero range

0

Off

1

On

A “0” setting implies: The initial zero setting mechanism (IZSM) will work

over the entire range of the scale capacity and the over-capacity point is 103%
above the zero obtained at calibration.

A “1” setting implies: The IZSM must be within +/- 10% of the zero

obtained at calibration and the over-capacity point is 103% above the IZSM.

19. The scale shows “EE-14

EE-14

EE-14

EE-14”. Press the zero key. The scale shows the current

graduation selection. Use the following table to select a graduation value.

Settings

Count by in lb

Count by in kg

0

100

50

1

50

20

2

20

10

3

10

5

4

5

2

5

2

1

6

1

0.5

7

0.5

0.2

8

0.2

0.1

9

0.1

0.05

10

0.05

0.02

11

0.02

0.01

12

0.01

0.01

NOTE: The stated accuracy specifications are based on the graduation setting in
the table below. If the graduation setting is set other than the value in the table
below the accuracy specification remains with the graduation size listed below.

If your capacity is:

Set your graduation to:

25 lb / 12.5 kg

.01 lb / .01 kg (12)

50 lb / 25 kg

.02 lb / .01 kg (11)

100 lb / 50 kg

.05 lb / .02 kg (10)

250 lb / 125 kg

.1 lb / 0.05 kg (9)

300 lb / 150 kg

.1 lb / 0.05 kg (9)

500 lb / 250 kg

0.2 lb / 0.1 kg (8)

1000 lb / 500 kg

0.5 lb / 0.2 kg (7)