User Defined Function Keys 13

Auto Total

The

Fx-TOTAL

key under the AUTO TOTAL mode functions as Auto Total On/ Auto

Total Off.
The Auto Mode has three variations which are programmed in the SETUP menu:

AutoLoad - Any settled tension above the ‘Rise above’ threshold will be

automatically totaled. Then the scale must fall below the ‘Drop below’ threshold

before another total is allowed.

AutoNorm - This mode takes the last settled tension to auto total with. The total

occurs only once the Scale goes below the threshold. This allows the load to be

adjusted without a total occurring. Once load is removed, the Scale uses the last

settled reading for total.

AutoHigh - Similar to the AutoNorm mode except the Scale uses the highest settled

reading. Useful for loads that can’t be removed all at once.

View Total

The

Fx-VIEW TOTAL

key activates the Total weight display followed by the number of

samples. While the display is showing the Total, Total is cleared by pressing

ZERO

.

Net/Gross

Switches the display between Net and Gross modes. Net Tension is defined as Gross

Tension minus a Tare Weight. To switch between Net Mode and Gross Mode press the

Fx-

NetGross

key (setup to the Net/Gross function).

The

Fx-NetGross

key will only function if a Tare value has been established.

Switching back to Gross mode from Net mode will not clear the Tare value. This allows

the operator to use the Gross Mode temporarily without having to reestablish the Tare

value. Only clearing the Tare or setting a new Tare will change the tare value held before

switching into Gross Mode.

Peak Hold

Peak Hold automatically updates the display when a higher peak tension reading is

established. The Peak Hold function uses a high speed mode of the A/D converter (220

samples) allowing it to capture transient tensions at a far higher rate than typical

Dynamometers. Peak hold is cleared and re-enabled with the

Fx-Peak Hold

Key. When a

new peak is detected, the Fx LED will flash three times. The accuracy of the system in

Peak Hold mode is slightly reduced to .2% of Capacity + 5d. The Filter setting is turned

off while in Peak Hold mode to ensure the fastest acquisition rate.

Example Peak Hold Application-
The Peak Hold function is useful in “Dynamic” and “Fall” tests. Common tests for fiber

rope include “Overall Breaking Strain” (OB€), Breaking Force, and Cycled Breaking

Strain. The MSI-8000 combined with a force test stand, meets the speed and accuracy

requirements to properly conduct these tests.