COMM Set Up 29

Serial Cable Schematic, DTE Configuration for connecting directly to a DCE printer or

other DCE devices (503489-0002 schematic).

5.10 Service Counters

All MSI’s RF Linked Scales and Dyna-Links maintain two service counters for safety. The

first counter (LFCnt) counts lifts above 25% of capacity. The second counter (OLCnt)

counts the number of times the RF Linked Scale has been overloaded. These counters

serve to warn the user to inspect the load train after a number of overloads or a long term

frequency of high capacity lifts. Service counters can only be reset by the factory. The

power up routine will be interrupted when the lift counter exceeds 16383 lifts or the

overload counter exceeds 1023 overloads. Push any key to continue operation.
Feature only available on MSI-8000 Software release 2.00 and above. Service counters are

available on the Scale/Dyna-Link front panel Test function.