CHAPTER 2 DIAGNOSIS

2.3 ERROR CODES

Scout Pro balances are equipped with software which will display an error condition when it occurs.

Table 2-2 Error Codes, describes the various error codes which can appear on the display and

specifies the probable reason and remedy.

TABLE 2-2. ERROR CODES

DISPLAY

REASON

ACTION

Err 1

Invalid checksum data

1. Perform factory calibration.

2. Restore data file.

3. Check transducer.

4. Check PCB.

Err 2

Overload or Underload

1. Make sure the right pan is on the balance.

2. Perform service LIN calibration per B.1.

3. Check transducer.

4. Check PCB.

Err 4

Incorrect calibration mass

1. Calibrate using correct mass.

Err 5

Parts counting error

1. Sample is less than 1d. Change sample quantity

or increase sample.

Err 8

RS232 buffer is full

1. Check printing device.

Err 9

Internal data error

1. Restore data file.

2. Check PCB.

3. Check transducer.

TABLE 2-1. DIAGNOSTIC GUIDE (Cont.)

SYMPTOM

RS232 Not functioning.

(On models equipped with

RS232).

PROBABLE CAUSE

1. Improper Print Menu set-

tings.

2. RS232 not turned on.

3. RS232 does not stay on.

4. Loose or disconnected

RS232 option.
5. RS232 option defective.

REMEDY

1. Check all settings. (Menu

item Power must be set

to ON.)
2. Turn the option on. (See

instruction manual).
3. If batteries are being

used, the RS232 turns

off when the unit is turned

off. Either turn the option

on or use the ac adapter.
4. Check connection.
5. Replace RS232 option.