4-5

CHAPTER 4 TESTING

TABLE 4-3. TEST MASS VALUES

Model

Mass 1

Mass 2

SPx202

250

350

SPx402

500

700

SPx401

500

900

SPx601

700

900

SPx2001

3000

7000

SPx4001

5000

7000

SPx6000

7000

9000

TABLE 4-4. TRANSDUCER ADJUSTMENT TOLERANCES

Model

Min(mm/in)

Max(mm/in)

SPx123

0.40/0.016

0.45/0/018

SPx202

0.60/0.024

0.70/0.028

SPx402

0.50/0/020

0.60/0.024

SPx401

0.75/0.030

0.90/0.035

SPx601

0.65/0.026

0.75/0.030

SPx2001

0.65/0.026

0.90/0.035

SPx4001

0.60/0.024

0.70/0.028

SPx6000

0.50/0/020

0.60/0.024

4.3 Down Stop Adjustment (Cont.)

After the transducer has been adjusted, assemble the balance retest and recalibrate.