3-3

CHAPTER 3 REPAIR PROCEDURES

3.1.2 Printed Circuit Board (9) Replacement (Cont.)

3. When an RS232 option is installed, remove the RS232 Connector from J11on the Printed

Circuit Board (9).

4. Disconnect the Transducer cable from Connector J6 on the Printed Circuit Board (9).

5. Disconnect the battery cable from Connector J2 on the Printed Circuit Board (9).

6. Remove the 2 mounting screws on top of the Printed Circuit Board (9).

7. Push the Printed Circuit Board towards the back as far as it will go and apply a small forward

pressure to the Cal Lock (15). Lift up on the front of the Printed Circuit Board (9) to remove

it.

8. Install the new Printed Circuit Board (9). Make sure that the Printed Circuit Board is

positioned under the 2 clips on the Base (10).

9. Connect the flexible cable from the Transducer to Connector J6 on the Printed Circuit

Board (9).

10.Connect the RS232 to Connector J11 on the Printed Circuit Board (9) if the RS232 is

installed.

11.Connect the battery cable to connector J2 on the Printed Circuit Board (9).

12.Replace the 2 mounting screws on top of the Printed Circuit Board (9), make sure the board

is positioned as far forward as possible to ensure the calibration lock switch will operate

properly.

13.Reassemble the Balance.

14.Replace the batteries if previously removed.

15.See Appendix A Service Tools Instructions and load temperature compensation data.

16.Check the performance of the Balance and perform Factory Calibration B.2.

Advertising