CHAPTER 4 TESTING

4.2.2 Off-Center Load Test

The Off-Center Load Test is used to determine whether displayed weight values are affected by

moving the sample to different areas of the Pan (6).

Test

1. Place 1/2 of the balance capacity in the center of the Pan (1).

2. Note the reading.

3. Move the mass halfway (between the center and the edge) to the front of the Pan (1). Note

any differences in the displayed weight reading.

4. Repeat this test for the back, left, and right positions of the Pan.

5. Maximum allowable change in displayed weight readings is + 2 count for each of the four

positions. If this reading is exceeded, it usually indicates a defective Transducer (8).

4.2.3 Linearity Test

This test is used to determine the linearity of the unit throughout its operating range. The masses used

to perform this test must be adjusted to ASTM Class 4 Tolerance, or better.

NOTE:

The balance must pass the Off-Center Load Test and Re-

peatability Test before the Linearity Test may be performed.

Test

Loads do not have to be test weights. They can be anything that totals the load value. The test mass

can be anything that weighs near the test mass value

TABLE 4-2. TEST MASSES

Capacity

Test mass

Load 1

Load 2

Load 3

Load 4

200g

1g

50g

100g

150g

190g

400g x 0.01

1g

100g

200g

300g

390g

400g x 0.1

5g

100g

200g

300g

400g

600g

5g

150g

300g

450g

600g

2000g

5g

500g

1000g

1500g

2000g

4000g

5g

1000g

2000g

3000g

4000g

6000g

10g

1500g

3000g

4500g

6000g

1. Place the test mass on the balance, record the weight and then remove.

2. Place Load 1 on the balance and press the ON/ZERO/Off button.

3. Place the test mass on the balance, record the weight and then remove.

4. Place Load 2 on the balance and press the ON/ZERO/Off button.