APPENDIX A SERVICE TOOL INSTRUCTIONS

2.

If the communications parameters have to be set, click on the Com Port tab and the screen

shown in Figure A-5 will appear. Adjust the parameters as required to match the balance.

Figure A-5.

A.5 USING CONFIGTOOL.EXE. (CONT.)

3.

When the communications parameters are matched click on the Test Command tab and

test the communications.

4.

In this example, see figure A-6, the “P” command resulted in the 0.02g being written in the

lower screen. The “V” command resulted in Sr: 1.10 being written. The T command had

the effect of zeroing the balance, and the last P command resulted in the 0.00g being

written. If the balance does not reliably communicate the problem must be fixed before

continuing.