Electromagnetic emissions

Voltage

fluctuations/flicker

emissions
IEC 61000-3-3

WARNING This equipment/system is intended for
use by healthcare professionals only. This

equipment/ system may cause radio interference or
may disrupt the operation of nearby equipment

a

. It

may be necessary to take mitigation measures, such

as re-orienting or relocating the monitor or shielding

the location.

Complies

a

The monitor contains a 5-GHz orthogonal frequency-division multiplexing transmitter or a 2.4-GHz frequency

hopping spread-spectrum transmitter for the purpose of wireless communication. The radio is operated

according to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/

EC). The transmitter is excluded from the EMC requirements of 60601-1-2, but should be considered when

addressing possible interference issues between this and other devices.

Electromagnetic immunity

The monitor is intended for use in the electromagnetic environment specified below. The customer or the

user of the monitor should assure that it is used in such an environment.

Immunity test

IEC 60601 test

level

Compliance

level

Electromagnetic environment - guidance

Electrostatic

discharge (ESD)
IEC 61000-4-2

±6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile.

If floors are covered with synthetic material, the

relative humidity should be at least 30%.

Electrical fast

transient/burst
IEC 61000-4-4

±2 kV for power

supply lines
±1 kV for input/

output lines

±2 kV for power

supply lines
±1 kV for input/

output lines

Mains power quality should be that of a typical

commercial or hospital environment.

Surge
IEC 61000-4-5

± 1 kV differential

mode
± 2 kV common

mode

± 1 kV differential

mode
± 2 kV common

mode

Mains power quality should be that of a typical

commercial or hospital environment.

Voltage dips, short

interruptions and

voltage variations

on power supply

input lines
IEC 61000-4-11

>95% dip in 0.5

cycle
60% dip in 5 cycles
30% dip for 25

cycles
>95% dip in 5

seconds

>95% dip in 0.5

cycle
60% dip in 5 cycles
30% dip for 25

cycles
>95% dip in 5

seconds

Mains power quality should be that of a typical

commercial or hospital environment. If the user

of the monitor requires continued operation

during power mains interruptions, it is

recommended that the monitor be powered from

an uninterruptible power supply or a battery.

120 Guidance and manufacturer's declaration