Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 16

Advertising
background image

WARNING Liquids can damage electronics inside the Connex IWS. Prevent

liquids from spilling on the wall system.

If liquids are spilled on the wall system:

1. Power down the wall system.

2. Disconnect the power plug.

3. Remove the wall system from the wall.

4. Remove battery pack from the wall system.

5. Dry off excess liquid from the wall system.

Note

If liquids possibly entered the wall system, remove the wall system

from use until it has been properly dried, inspected, and tested by

qualified service personnel.

6. Reinstall battery pack.

7. Mount the wall system on the wall.

8. Power on the wall system and verify that it functions normally before using it.

WARNING Safety risk. Damaged cords, cables, and accessories can affect

patient and operator safety. Routinely inspect the AC power cord, blood pressure

cuff, SpO2 cable, and other accessories for strain relief wear, fraying, or other

damage. Replace as necessary.

WARNING Fire and explosion hazard. Do not operate the monitor in the presence

of a flammable anesthetic mixture with air, oxygen, or nitrous oxide; in oxygen-

enriched environments; or in any other potentially explosive environment.

WARNING Fire and shock hazard. Only connect LAN cables contained within the

perimeter of a single building. Conductive LAN cables spanning multiple buildings

may introduce fire or shock hazards unless they are fitted with fiber optic cables,

lightning arrestors, or other applicable safety features.

WARNING The monitor may not function properly if dropped or damaged. Protect

it from severe impact and shock. Do not use the monitor if you notice any signs of

damage. Qualified service personnel must check any monitor that is dropped or

damaged for proper operation before putting the monitor back into use.

WARNING Defective batteries can damage the monitor. If the battery shows any

signs of damage or cracking, it must be replaced immediately and only with a

battery approved by Welch Allyn.

WARNING Improper disposal of batteries may create an explosion or

contamination hazard. Never dispose of batteries in refuse containers. Always

recycle batteries according to local regulations.

WARNING Electric shock hazard. Do not open the monitor or attempt repairs. The

monitor has no user-serviceable internal parts. Only perform routine cleaning and

maintenance procedures specifically described in this manual. Inspection and

servicing of internal parts shall only be performed by qualified service personnel.

WARNING Inaccurate measurement risk. Do not expose to temperatures higher

than 122º F (50º C).

WARNING Inaccurate measurement risk. Do not use the monitor on patients who

are on heart-lung machines.

WARNING Use the monitor only as described in this directions for use. Do not

use the monitor on patients as described in the Contraindications.

10 About warnings and cautions

Welch Allyn Connex

®

Integrated Wall System