Electromagnetic immunity

The monitor is intended for use in the electromagnetic environment specified below. The customer or the

user of the monitor should assure that it is used in such an environment.

Immunity test

IEC 60601 test

level

Compliance

level

Electromagnetic environment - guidance

Portable and mobile RF communications

equipment should be used no closer to any part

of the monitor, including cables, than the

recommended separation distance calculated

from the equation applicable to the frequency of

the transmitter.

Recommended separation distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz

3 Vrms

d= (1.17)

Radiated RF
IEC 61000-4-3

3 V/m 80 MHz to 1

GHz

3 V/m

d= (1.17)

80 to 800 MHz

d= (2.33)

800 MHz to 2,5 GHz

whereP is the maximum output power rating of

the transmitter in watts (W) andd is the

recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey

a

,

should be less than the compliance level in each

frequency range

b

. Interference may occur in the

vicinity of equipment marked with the following

symbol:

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects, and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted

theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an

electromagnetic site survey should be considered. If the measured field strength in the location in which the

monitor is used exceeds the applicable RF compliance level above, the monitor should be observed to verify

normal operation. If abnormal performance is observed, additional measures may be necessary, such as

reorienting or relocating the monitor.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Directions for use

Guidance and manufacturer's declaration 121