If

Display MAP

is selected, specify which

numerics are primary in the NIBP frame. On the

Home tab, clinicians can touch the NIBP frame to

toggle between views.

Default patient type

Select a default patient type for this monitor. The

patient type shows in the Patient frame on the

Home tab.
In the Patients tab on the Summary tab, clinicians

can change the displayed patient type from the

default patient type that you set here.

Tube type

Select the number of tubes that are connected to

the NIBP cuff that is used with this monitor. If you

select

1 tube

, the only algorithm available for

selection is

Step

.

Unit of measure

Select the NIBP unit of measure for display.

Allow interval program changes

Enable clinicians to modify interval program

settings from the Intervals tab.

Algorithm and Cuff inflation target (CIT)

Select the default algorithm used to determine

NIBP measurements.

If you select the

Step

algorithm, touch

and

enter a default cuff inflation target for each type of

patient. In the Patients tab on the Summary tab,

clinicians can change the CITs from the default

CITs that you set here.

5. Do one of the following:

To continue in the Advanced Settings, touch another tab.

To exit the Advanced Settings and return to the Home tab, touch Exit.

Specify advanced temperature settings

1. Access the Advanced Settings.

a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab will appear.

2. Touch the Parameters tab.
3. Touch the Temperature tab.
4. Specify settings.

Setting

Action/Description

Unit of measure

Select primary units of measure for the

temperature display on the Home tab

Display temperature conversion

Select this to display primary units of measure and

secondary units of measure for the temperature

display on the Home tab.

Default SureTemp Plus site

Select the default site for SureTemp

measurements. The default site applies when

clinicians power up the monitor and each time