Take a temperature at the rectal site – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 90

Advertising
background image

Note

The monitor does not retain Direct mode temperatures in memory.

Therefore, it is important to note the temperature before removing

the probe from the measurement site and then manually record it in

the patient record.

6. Remove the probe after the temperature measurement is complete and firmly press the eject

button on the top of the probe to release the probe cover.

7. Return the probe to the probe well to continue taking temperatures in the Predictive mode.
8. Wash your hands to reduce the risk of cross-contamination.

Take a temperature at the rectal site

WARNING Patient injury risk. When taking rectal temperatures, insert the probe

tip only 5/8 inch (approximately 1.5 cm) inside the rectum of adults and only 3/8 inch

(approximately 1 cm) inside the rectum of children to avoid the risk of bowel

perforation.

WARNING Cross-contamination or nosocomial infection risk. Thorough hand-

washing greatly reduces the risk of cross-contamination and nosocomial infection.

WARNING Patient injury risk. Do not exceed the recommended temperature

measurement durations in Direct mode. Continuous measurement durations of 3

minutes at the oral and rectal sites and 5 minutes at the axillary site are

recommended for accurate measurement. Do not continuously measure beyond 10

minutes in any mode.

WARNING Inaccurate measurement risk. To ensure optimal accuracy, always

confirm that the correct mode and site are selected.

CAUTION Probe covers are disposable, nonsterilized, and single-use. Probes are

also nonsterilized. Do not autoclave probes and probe covers. Ensure that probe

covers are disposed of according to facility requirements or local regulations.

1. Remove the rectal temperature probe from the rectal probe well.

The monitor sounds a tone as it enters the ready state. The Temperature Site Control defaults

to the rectal site.

2. Insert the rectal probe into a new probe cover and press the probe handle down firmly.
3. Separate the patient’s buttocks with one hand. Use the other hand to gently insert the probe

tip only 5/8 inch (1.5 cm) inside the rectum of adults and only 3/8 inch (approximately 1 cm)

inside the rectum of children. The use of a lubricant is optional.

4. Insert the probe so that the tip is in contact with tissue. Continue to separate the buttocks and

hold the probe in place throughout the measurement process. While the measurement is

taking place, the temperature frame displays the process indicator.

84 Patient monitoring