Recommended separation distances between portable and mobile RF
communications equipment and the monitor

The monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are

controlled. The customer or user of the monitor can help prevent electromagnetic interference by

maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)

and the monitor as recommended below, according to the maximum output power of the communications

equipment.

Separation distance according to frequency of transmitter (m)

Rated max.

output power

of transmitter

(W)

150 kHz to 80

MHz
d= (1.17)

80 MHz to 800

MHz
d= (1.17)

800 MHz to 2.5 GHz
d= (2.23)

0.01

0.11667

0.11667

0.23333

0.1

0.36894

0.36894

0.73785

1

1.1667

1.1667

2.3333

10

3.6894

3.6894

7.3785

100

11.667

11.667

23.3333

For transmitters rated at a maximum output power not listed above, the recommended separation distanced

in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, whereP is

the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects, and people.

122 Guidance and manufacturer's declaration