Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 96

Advertising
background image

WARNING Inaccurate measurement risk. Use only Nellcor sensors and

accessories on Nellcor-equipped monitors.

WARNING The pulsations from intra-aortic balloon support can increase the

pulse rate displayed on the monitor. Verify the patient's pulse rate against the ECG

heart rate.

WARNING Patient injury risk. Do not attempt to reprocess, recondition, or recycle

any sensors or patient cables. Doing so might damage electrical components.

WARNING Pulse rate measurement might not detect certain arrhythmias

because it is based on the optical detection of a peripheral flow pulse. Do not use

the pulse oximeter as a replacement or substitute for ECG-based arrhythmia

analysis.

WARNING Use the pulse co-oximeter as an early warning device. As you

observe a trend toward patient hypoxemia, use laboratory instruments to analyze

blood samples to better understand the patient's condition.

WARNING The accuracy of SpO2 measurements can be affected by any of the

following:

elevated levels of total bilirubin

elevated levels of Methemoglobin (MetHb)

elevated levels of Carboxyhemoglobin (COHb)

hemoglobin synthesis disorders

low perfusion at the monitored site

the presence of concentrations of some intravascular dyes, sufficient to change the

patient's usual arterial pigmentation

patient movement

patient conditions such as shivering and smoke inhalation

motion artifact

painted nails

poor oxygen perfusion

hypotension or hypertension

severe vasoconstriction

shock or cardiac arrest

venous pulsations or sudden and significant changes in pulse rate

proximity to an MRI environment

moisture in the sensor

excessive ambient light, especially fluorescent

the use of the wrong sensor

a sensor applied too tightly

1. Verify that the sensor cable is connected to the monitor.

90 Patient monitoring

Welch Allyn Connex

®

Integrated Wall System

Advertising