Note

In the Continuous Monitoring profile, all patient measurements remain
onscreen when automatic interval measurements are saved.

The

button changes to a timer (

), which counts down to the next

automatic measurement.

Automatic measurements continue until you turn off intervals.

WARNING Patient harm risk. Do not use intervals on neonates out of
earshot. Verify that audio can be heard from where you intend to be.

Set up NIBP intervals

Follow these steps to configure NIBP intervals.

1. Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the Intervals vertical tab.
4. Select Automatic, Program, or Stat, and follow the steps presented in the

associated topics that follow.

5. To automatically print patient data at each interval, touch the Automatic print on

interval checkbox.

6. To start intervals immediately, touch Start intervals. Otherwise, touch the Home

tab.
The new settings take effect immediately.

Automatic intervals

You can configure the monitor to take automatic NIBP measurements at consistent

intervals. Use the spin box or keypad to set the desired interval.

Note

An alarm does not turn off intervals. Subsequent automatic measurements
continue to occur as scheduled.

128 Patient monitoring