Masimo specifications (for additional information, refer to the directions for use that

accompanies each accessory)

Adults, Infants, Pediatrics, Neonates (low perfusion

4

): ±2%

Resolution

1%

Perfusion Index

Measurement range

0.1% to 20.0%

Pulse rate

Unit of measure

beats per minute

Measurement range

25 to 240 beats per minute

Accuracy

5

Adults, Infants, Pediatrics, Neonates (no motion): ±3 beats

per minute
Adults, Infants, Pediatrics, Neonates (motion): ±5 beats per

minute
Adults, Infants, Pediatrics, Neonates (low perfusion

4

): ±3

beats per minute

Resolution

1 beat per minute

SpHb

Units of measure

g/dL, mmol/L; user-selectable

Measurement range

0.0 to 25.0 g/dL (0.0 to 15.5 mmol/L)

Accuracy

6

Adults, Pediatrics (no motion): 8 to 17 g/dL ±1 g/dL

Resolution

0.1 g/dL or mmol/L

RRa

Unit of measure

breaths per minute

Body weight

Adult >66 pounds (30 kilograms)

Measurement range

0 to 70 breaths per minute

Accuracy

7

Adults: 4 to 70 ±1 breath per minute

Resolution

1 breath per minute

1

SpO2 accuracy was determined by testing on healthy adult volunteers in the range of 60% to 100% SpO2 against a

laboratory CO-Oximeter. SpO2 accuracy was determined on 16 neonatal NICU patients ranging in age from 7 to 135 days

old and weighing between 0.5 and 4.25 kg. Seventy-nine (79) data samples were collected over a range of 70% to 100%

SaO2 with a resultant accuracy of 2.9% SpO2. Contact Masimo for testing specifications.

2

The Masimo sensors have been validated for no-motion accuracy in human blood studies on healthy adult male and

female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70% to 100% SpO2

Directions for use

Specifications 201