Resume continuous monitoring

You can resume continuous monitoring after a pause when patient sensors are

connected to a patient. The status of the pause timer determines your next step.

Pause time remaining

If the pause time limit has not expired (time remains on the countdown timer), resume

monitoring as follows:

1. Reconnect patient sensors to the patient as needed.
2. Touch Resume monitoring.

The Home tab appears and continuous monitoring resumes.

Pause time limit expired (patient sensors connected)

If the pause time limit has expired and you have already reconnected patient sensors to

the patient and/or the device, the Home tab appears and continuous monitoring

automatically resumes.

Pause time limit expired (patient sensors disconnected)

If the pause time limit runs out before you resume monitoring, the device exits Pause

mode, and alarms might occur.

Reconnect patient sensors to the patient and/or device and acknowledge alarm and

information messages as needed.
Continuous monitoring resumes.

End continuous monitoring

When you no longer need to monitor a patient continuously, follow these steps to end

monitoring:

1. Touch the Settings tab.

The Monitor tab appears.

2. Touch End monitoring.

An "End monitoring" dialog appears with the following options:

New patient – Clears patient and location context and review data after you

remove patient sensors

Power down – Clears patient and location context and review data, and powers

down the device

Cancel – Retains patient data and returns to Home tab

3. Select the desired option.