Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual

Page 219

Advertising
background image

CO2 specifications (for additional information, refer to the directions for use that accompanies

each accessory)

Initialization time

40 seconds (typical, includes power-up and initialization

time)

System response time

3.2 seconds (typical, includes module response time and

host monitor system response time)

Compensation

The CO2 module is equipped with a barometric pressure

transducer and compensation is triggered at startup or

during other events (significant changes in temperature,

ambient pressure, etc.).
BTPS (standard correction used by Microstream capnography

during all measurement procedures for body temperature,

pressure, and saturation).

Cyclical pressure

10 kPa (100 cmH

2

O); module operates within specification

with overpressure up to 100 cmH

2

O

Sampled gas return/disposal

Sampled gases are not returned to the breathing circuit.

Exhaled gas is exhausted from the monitor's exhaust port.

Dispose of sampled gases according to facility requirements

or local regulations.

Calibration interval

Initial: After 1200 operating hours
Subsequent: After 4000 operating hours or annually

(whichever comes first)

Periodic service

After 30,000 operating hours

etCO2

Units of measure

mmHg, kPa; user-selectable

Display range

0 to 150 mmHg (0.0 to 20.0 kPa)

Resolution

1 mmHg, 0.1 kPa

FiCO2

Units of measure

mmHg, kPa; user-selectable

Display range

0 to 150 mmHg (0.0 to 20.0 kPa)

Resolution

1 mmHg, 0.1 kPa

IPI

Display range

1 to 10

RR

Unit of measure

breaths per minute

Display range

0 to 150 breaths per minute

Directions for use

Specifications 213