Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 219
CO2 specifications (for additional information, refer to the directions for use that accompanies
each accessory)
Initialization time
40 seconds (typical, includes power-up and initialization
time)
System response time
3.2 seconds (typical, includes module response time and
host monitor system response time)
Compensation
The CO2 module is equipped with a barometric pressure
transducer and compensation is triggered at startup or
during other events (significant changes in temperature,
ambient pressure, etc.).
BTPS (standard correction used by Microstream capnography
during all measurement procedures for body temperature,
pressure, and saturation).
Cyclical pressure
≤
10 kPa (100 cmH
2
O); module operates within specification
with overpressure up to 100 cmH
2
O
Sampled gas return/disposal
Sampled gases are not returned to the breathing circuit.
Exhaled gas is exhausted from the monitor's exhaust port.
Dispose of sampled gases according to facility requirements
or local regulations.
Calibration interval
Initial: After 1200 operating hours
Subsequent: After 4000 operating hours or annually
(whichever comes first)
Periodic service
After 30,000 operating hours
etCO2
Units of measure
mmHg, kPa; user-selectable
Display range
0 to 150 mmHg (0.0 to 20.0 kPa)
Resolution
1 mmHg, 0.1 kPa
FiCO2
Units of measure
mmHg, kPa; user-selectable
Display range
0 to 150 mmHg (0.0 to 20.0 kPa)
Resolution
1 mmHg, 0.1 kPa
IPI
Display range
1 to 10
RR
Unit of measure
breaths per minute
Display range
0 to 150 breaths per minute
Directions for use
Specifications 213