WARNING Inaccurate measurement risk. The pulse co-oximeter can be
used during defibrillation, but the reading may be inaccurate for up to 20
seconds.

WARNING Patient injury risk. Do not attempt to reprocess, recondition, or
modify any sensors or patient cables. Doing so might damage electrical
components.

WARNING Pulse rate measurement might not detect certain arrhythmias
because it is based on the optical detection of a peripheral flow pulse. Do
not use the pulse oximeter as a replacement or substitute for ECG-based
arrhythmia analysis.

WARNING Use the pulse co-oximeter as an early warning device. As you
observe a trend toward patient hypoxemia, use laboratory instruments to
analyze blood samples to better understand the patient's condition.

WARNING The accuracy of SpHb measurements can be affected by any
of the following:

elevated levels of total bilirubin

elevated levels of Methemoglobin (MetHb)

elevated levels of Carboxyhemoglobin (COHb)

hemoglobin synthesis disorders

low perfusion at the monitored site

the presence of concentrations of some intravascular dyes, sufficient to

change the patient's usual arterial pigmentation

patient movement

patient conditions such as shivering and smoke inhalation

motion artifact

painted nails

poor oxygen perfusion

hypotension or hypertension

severe vasoconstriction

shock or cardiac arrest

venous pulsations or sudden and significant changes in pulse rate

proximity to an MRI environment

moisture in the sensor

excessive ambient light, especially fluorescent

the use of the wrong sensor

a sensor misapplied or dislodged

severe anemia

venous congestion

1. Verify that the sensor cable is connected to the monitor.

Directions for use

Patient monitoring 151