Troubleshooting and maintenance

Welch Allyn AcuityLink Clinician Notifier

Alarm notification does not seem to be
reliable.

Warning: Preserve Clinician Notifier program files and maintain
mobile device settings that were configured during installation of
AcuityLink Clinician Notifier software. Modifying, deleting or
replacing any Clinician Notifier files or changing any of the following
mobile device settings can result in sub-optimal Clinician Notifier
performance, including inhibition of patient alarm delivery: alarm
clock, backlight, external power, power off, screen alignment and
orientation, time settings, volume, wake-up and wireless network
connection.

Connecting mobile devices to the Internet, or running too many
programs in the foreground or background while the Clinician Notifier
program is running, can greatly reduce the battery life and memory of
the devices, thereby compromising alarm delivery.

Non-Clinician Notifier audio or vibratory notification, barcoding,
passwords or popup windows can interfere with effective Clinician
Notifier program performance, including alarm notification.

An alarm popup window interrupted
my logout sequence, and now I cannot
log out.

If you tap within an alarm popup window after you have tapped OK to log out,
tap

until the Logging Out popup window re-appears. To log out, tap Yes.