Directions for Use

View patient information

27

To respond to an alarm

1.

Tap the alarm listing.

2.

Tap one of these buttons:

Tap Suspend to suspend the alarm.

The Suspend button changes to Resume. If you want to resume the alarm, tap
Resume.

Notification of all alarms for this patient is suspended for 90 seconds on all mobile
devices that received the alarm. However, new lethal arrhythmia alarms and
apnea alarms for the patient interrupt the suspend state.

Tones are also suspended for this patient at the Acuity Central Station and the
patient monitor.

Tap Busy to indicate that you are busy.

Notification for this alarm stops on your device only.

If someone suspends the alarm and it remains active after the 90 second period,
notification returns to your device.

To navigate to another page for more information

1.

Tap the alarm listing in which you are interested.

2.

Tap one of these buttons.

Tap Details to view details of an alarm.

The Details page appears.

Tap Realtime to view a patient’s realtime waveform.

The Realtime page appears.

If Realtime is grayed out, the patient is not connected to the Acuity System.

In both of these pages, if the patient is in alarm, you can suspend the alarm.