Figure 5.4: Drain System Information

This machine must be installed, adjusted, and serviced by a qualified electrical
maintenance personnel familiar with the construction and operation of this type
of machinery. They must also be familiar with the potential hazards involved. If
this warning is not observed, personal injury or equipment damage resulting in
voiding the warranty may result.

transformer kit available from the manufacturer. Voltages above 250V and 490V require additional
measures. Contact the distributor or the manufacturer for assistance.

Table 5.3: Electrical Service Requirements

200-240V

380-480V

Machine

Max Amps

Breaker

Wire Size

Max Amps

Breaker

Wire Size

SI-110

21

30

10ga / 4mm

11

15

10ga / 4mm

SI-135

32

40

8ga / 10mm

16

20

10ga / 4mm

SI-200

58

60

6ga / 16mm

29

30

8ga / 10mm

SI-275

79

80

6ga / 16mm

39

40

8ga / 10mm

SI-300

89

90

6ga / 16mm

45

50

8ga / 10mm

Note: Do not use phase adders (roto-phase) on inverter driven equipment!

Note: Wire sizes shown are for copper, THHN, 90 conductor per NEC article 310 (USA).

The machine should be connected to an individual branch circuit not shared with lighting or other

If controlling the AC drive with a parameter unit, the machines computer and
its safety features are bypassed. This could allow the basket to rotate at high
speeds with the door open. When using a parameter unit to control the AC
drive, a large sign should be placed on the front of the machine warning people
of the imminent danger.