5 Putting into operation

24

KjelFlex K-360 Operation Manual, Version C

5 .3

Reagent and water connections

1

2

3

6

7

5

4

a H

3

BO

3

inlet from tank

b H

2

O inlet from tank

c NaOH inlet from tank
d Reagent inlet for acid resistant pump
e Cooling water inlet
f Cooling water outlet
g Drain to waste

Fig. 5.2: Reagent and water connections

All pumps are self-priming, no overpressure is necessary at the tanks.

5 .3 .1

Cooling water connection

The water pressure should be at a maximum of 4 bar. The built in valve reduces the water flow to 1.2
liters per minute.

The flanged screw coupling for the water connection has a standard screw thread of G ¾”.

5 .3 .2

Drainage of cooling water

Place the drain hose for the cooling water directly into the drain. For this purpose, shorten the silicone
hose to the optimal length. The drain hose should not show any kinks, sharp bends and/or siphoning
effect. Prevent flooding inside and outside the instrument by securing the drain hose.