7 Maintenance

50

KjelFlex K-360 Operation Manual, Version C

NOTICE

Risk of sealing damage by sharp edges or lubricants.

When installing, move sealing in perpendicular axis to the glass parts.

Do not touch sealing with potentially sharp objects.

Do not apply grease or any other lubricant to the sealing.

For further information on how to unmount and remount the splash protector, see chapter 7.3.3.

7 .3 .2

Replacing the splash protector

Replace the Splash protector after approximately 5000 determinations.

7 .3 .3

Replacing the rubber bung seals and the splash protector

We recommend to replace the parts listed below depending on the instrument use, at least once a
year.

The seals are subject to wear and tear, thus you should check them regularly. We recommend to
replace them once a year. The following parts need to be replaced (see also the spare parts list):

1

2

3

a Accessories splash protector
b Gasket for splash protector, complete
c SVL 22 sealing

Make sure that the instrument is cooled down.
Open the service door and loosen the two tubings
shown on the picture.