7 Maintenance

52

KjelFlex K-360 Operation Manual, Version C

Pull out the steam tubing.

To replace the rubber parts pull off the connecting
screw. The smaller sealing is located inside.

Fig. 7.1: Installing the glass assembly

Exchange all the rubber and plastic parts shown on the first picture of chapter 7.3.3.
Put back the replaced parts in the reverse order and make sure to tighten the screw on top of the
splash protector very carefully with the Polygrip.

7 .3 .4

Glass parts

Replace the sample tubes and the condenser if broken.

7 .3 .5

Replacing the hoses

Depending on the applications and the used chemicals, it might be necessary to replace some of the
hoses inside of the instrument to avoid leakages or aspiration problems.