8 Troubleshooting

56

KjelFlex K-360 Operation Manual, Version C

Table 8-1: General malfunctions and their remedy (cont.)

Error number / Error indication

Possible cause

Corrective measure

59

EEPROM, CPU

Contact the Buchi customer service

60

You cannot change the Standard Method!

Store the new parameter under a new
method name

61

You cannot delete the Standard Method!

Do not try to delete the standard method

62

You have changed an existing method.
Overwrite method?

Overwrite the existing method or store the
new parameter under a new method name

63

The method list is full! You have to delete
some methods.

Delete some methods and add new
methods

64

Restart from Watchdog!

No corrective measure needed!

65

You changed between the demo and the
normal mode.

Restart the system to activate the selected
mode

66

Attention: Reposition stop-cock! (Reagent)

Check the correct position of the stop-cock

67

Attention: Reposition stop-cock! (NaOH)

Check the correct position of the stop-cock

8 .2

Diagnostics

The K-360 software enables to carry out some operational, sensor, service and printer test and to
view some hardware information and the operating hours.

Fig. 8.1: Diagnostics

8 .2 .1

Operational Tests

This submenu enables to test the following components:

Valves

Pumps

Stirrer

Alarm

Steam Unit Heating

Titrator Out (Titrator suitable)

Dispenser Out (Dispenser suitable)

Battery Monitor