8 Troubleshooting

57

KjelFlex K-360 Operation Manual, Version C

To carry out an Operational Test proceed as follows:

Set the system offline by pressing the corresponding button.

Go to the part you want to test, e.g. Steam valve and switch it on by pressing the forward or the

backward button.
If the valve works properly, a press sound is audible.

Switch the valve off.

Set the system online again.

NOTE
The button „Sensors“ enables to switch to the submenu Test Sensors.

8 .2 .2

Test sensors

This submenu enables to test the sensor functionality.
To test e.g. the tube shield sensor, proceed as follows:

Open the tube shield. When the sensor is working properly “Open” will be indicated on the display.

Close the tube shield. Now “Close” has to be indicated on the display.

NOTE
The button „Actors“ enables to switch to the submenu Operational Tests.

8 .2 .3

Test Printer

This submenu enables to test the functionality of the printer by e.g. printing a test page.

8 .2 .4

Operating Hours

This submenu indicates the operating hours of the steam generator, the number of distillations and the
date and time of the last reset.

8 .2 .5

Hardware Info

This submenu indicates some hardware information, e.g. part and serial numbers and the production
date of the instrument.

8 .2 .6

Service Test

This submenu enables the instrument manufacturers or service technicians to carry out a functionality
test of the instrument.