4 .5 .4

Extractions with rinsing and with drying

Step 1: Enter parameters of step 1 according to table.
Step 2: Enter lower heating level and rinsing time.
Step 3-9: Enter lower heating and drying time for every further step. Store program.
Table with parameters to be entered:

Soxhlet Standard

Soxhlet Warm

Hot Extraction

Continuous

Step 1

lower heating
level

Extraction time
and / or number
of cycles

upper and lower
heating level

Extraction time
and / or number of
cycles

upper and lower
heating level

Extraction time

lower heating
level

Extraction time

Step 2

lower heating level

rinsing time

Step 3 - 9

lower heating level

drying time

The drying can be performed by decreasing the lower heating level from step to step (3-9). The
sample is dried slowly. This is recommended for volatile and/or heat-sensitive analytes.
A mild drying is performed by using the inert gas option to prevent oxidation. It is activated by pressing
the INERT GAS key and remains activated until the step is finished. When the program is expired the
solvent is collected in the extraction chamber (automated solvent recovery) and the analyte/extract is
in the beaker.