7 Maintenance and repairs

45

B-811 Operation Manual, Version G

7 .4

Magnetic glass valve

5

1

3

4

2

6

a Cover cap
b Glass valve
c Valve housing
d Anchor
e Spring
f Inductor coil

Fig. 7.1: Exploded view of the glass valve

Insert the spring

e into the anchor d.

Screw the valve housing

c and the magnetic valve together f.

Push the valve onto the glass part

b.

Screw the cover

a on and tighten slightly.

7 .5

Membrane assembly

Insert anchor with membrane at an angle into

valve housing
Guide entirely through opening by pressing

slightly with circular movements.

Fig. 7.2: Membrane assembly

7 .6

Heating elements

The heating elements should be cleaned regularly after they have cooled down. Remove dirt adhering
to the plate with e.g. ethanol. Then clean the plate with a moist cloth and dry it.

ATTENTION
Never use scratching sponges, cleaning pads or cleaner containing abrasives to clean the heating
elements as they might damage the surface.