5 Putting into operation

34

B-811 Operation Manual, Version G

5 .7 .4

Connection to the condensers

1

2

3

2

2

2

4

left

right

a Left condenser connection

to impeller water meter and
then to chiller inlet

b Not used
c Water connection between

the condensers

d Right condenser connec-

tion to chiller outlet

Fig. 5.6: Connection to the condensers (rear view)

To connect the condensers proceed as follows:

Untie the tubes

b and fix them with cable binders.

Connect a silicone tube

c (approx. 35 cm) by means of 2 tube adapters to the condensers in the

middle.
For water inlet connect a silicone tube with adapters to the right condenser

d and connect the

adapters to the chiller outlet.
Connect the silicone tube (approx. 90 cm) from the rear side of the instrument

a (in Fig. 5.5) to the

tube adapter of the left condenser a (in Fig. 5.6).