8troubleshooting, 8 .1 malfunctions and their remedy – BUCHI Extraction System B-811 LSV User Manual

Page 48

Advertising
background image

8 Troubleshooting

48

B-811 Operation Manual, Version G

8

Troubleshooting

This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a Buchi technical
engineer who has access to the official service manuals. In this case, please refer to your local Buchi
customer service agent.

8 .1

Malfunctions and their remedy

Table 8-1: General malfunctions and their remedy

Error number

Possible cause

Corrective measure

Error 1

Empty beaker

Stop the program and refill the solvent beaker

Error 2

Beaker missing

Insert the beaker and turn on the heating
element

Error 3

Optical sensor error

Contact the Buchi customer service

Error 4

Draining not possible

Clean the glass valve

Error 5

Evaporating time > 30 min
between the heating elements

Heating power too low, volume differences are
too large

Error 10

No cooling water

Open the cooling water flow

Error 11

No location active

Insert a beaker and turn on the heating element

Technical errors:
Error 30

Temperatures too low on 1. heating
element

Error 31

Temperatures too high, lower
heating

Error 32

Lower heating circle broken

Error 33

Temperatures too low on upper
heating element

Error 34

Temperatures too high on upper
heating element

Error 35

Upper heating circle broken

Other error messages

Technical error

Contact the Buchi customer service