7 Maintenance and repairs

46

B-811 Operation Manual, Version G

7 .7

Holder ring with seal

2

1

a Holder ring
b PTFE seal

Fig. 7.3: Holder ring with seal

ATTENTION
When removing and reinstalling the seals, make sure not to damage them.

Never apply grease to the seals and never touch them with sharp objects, otherwise they will get
damaged and cause solvent loss.

To prolong the lifetime of the seals, rinse them regularly with distilled water or alcohol. Afterwards, dry
them with a soft cloth.

7 .8

Glass sample tubes

The glass sample tubes can be cleaned using commercial cleaning agent or in a dishwasher.

7 .9

Solvent beakers

The roughened area at the bottom of the solvent beakers prevents solvents from bumping. In order to
maintain this area in active state the beakers must be cleaned properly after every extraction.

7 .10

Emergency release mechanism

> 80 °

In case of power outage the samples can be
removed by unlocking the lift with a screwdriver.
The screwdriver is inserted at an angle of about
80° into the square slit on top of the instrument.
The mechanism is released by slightly tipping
the screwdriver downwards.

Fig. 7.4: Emergency release mechanism