Function key

Display

Description

Display

STEP 2 blinks (STEP 2 = Rinsing) Only the lower heating elements are activated

Display

HEATING

blinks

Select the heating level (1 - 20) using the / keys
Select level 13 for toluene
In case of oxygen-sensitive analytes, the gas supply must be
activated by pressing the INERT GAS key
The inert gas valve is automatically opened up while the step
starts

Display

H: MIN blinks Select the rinse time (0 to 99:99 h) using the / keys

The only stop criterion used for

STEP 2 is the rinse time

Display

STEP 3 blinks (STEP 3 = Drying)

Display

HEATING

blinks

Select the heating level (1 - 20) using the / keys
Select level 4 for toluene

Display

H: MIN blinks Select the drying time (0 to 99:99 h) using the / keys

The only stop criterion used for

STEP 3 is the drying time

In case of oxygen-sensitive analytes, the gas supply must be
activated by pressing the INERT GAS key
The inert gas valve is automatically opened up while the step
starts

SELECT No display blinks

The entered program is stored as program 0
To store this program under a different program number, please
see chapter 6.7

6 .5

Hot Extraction Mode - programming example

Program description:
This program carries out a Hot Extraction according to Randall and Goldfisch. In case of a Hot Extrac-
tion the sample is placed into the boiling solvent and extracted. The optical sensor detects the solvent
level in the extraction chamber, the magnetic valve is opened and a few milimeter of solvent are
drained. At the same time fresh solvent is constantly condensing at the condenser and collected in the
chamber. Solvent: toluene, for other heating levels see chapter 6.11.

Function key

Display

Description

SELECT The Mode display

blinks

Select the

Hot Extraction Mode using the / keys

Display

STEP 1 blinks (STEP 1 = Extraction)

Display

HEATING

blinks

(The symbol for the

upper heating element lights up)

Select the heating level (1 - 10) using the / keys
Select level 6 for toluene

Display

HEATING

blinks

(The symbol for the

lower heating element lights up)

Select the heating level (1 - 20) using the / keys
Select level 13 for toluene

Advertising