6 .4 soxhlet warm mode - programming example – BUCHI Extraction System B-811 LSV User Manual

Page 37

Advertising
background image

6 Operation

37

B-811 Operation Manual, Version G

Function key

Display

Description

Display

H: MIN blinks Select the rinse time (0 to 99:99 h) using the / keys

The only stop criterion used for

STEP 2 is the rinse time

Display

STEP 3 blinks (STEP 3 = Drying)

Display

HEATING

blinks

Select the heating level (1 - 20) using the / keys
Select level 4 for petroleum ether 40/60

Display

H: MIN blinks Select the drying time (0 to 99:99 h) using the / keys

The only stop criterion used for

STEP 3 is the drying time

In case of oxygen-sensitive analytes, the gas supply must be
activated by pressing the INERT GAS key
The inert gas valve is automatically opened up while the step
starts

SELECT No display blinks

The entered program is stored as program 0
To store this program under a different program number, please
see chapter 6.7

6 .4

Soxhlet Warm Mode - programming example

Program description:
This program is the same as the program for the Soxhlet Standard mode, except that the extraction
chambers are heated up. This makes it possible to reduce the extraction time, especially when using
solvents with high boiling points. Solvent: toluene, for other heating levels see chapter 6.11.

Function key

Display

Description

SELECT The Mode display

blinks

Select the

Soxhlet Warm Mode using the / keys

Display

STEP 1 blinks (STEP 1 = Extraction)

Display

HEATING

blinks

(The symbol for the

upper heating element lights up)

Select the heating level (1 - 10) using the / keys
Select level 6 for toluene

Display

HEATING

blinks

(The symbol for the

lower heating element lights up)

Select the heating level (1 - 20) using the / keys
Select level 13 for toluene

Display

CYCLES

blinks

Select the number of cycles using the / keys
If the number of cycles = 0 is entered, then the extraction is
finished when the extraction time has expired

Display

H: MIN blinks Select the extraction time (0 - 99:99 h) using the / keys

If no time is entered, then the extraction is finished when all posi-
tions have completed the set number of cycles
If both parameters are entered (time and number of cycles), then
both criteria must be fulfilled before the instrument switches to
the next step

Advertising