The Calibration window

Compare this value to your archive of the calibration constant data (if
available) for bulb 1. The values should be identical.

3. If they are not, place your cursor in the appropriate field, delete the

entry, then type the correct values for the constant in the text box.

4. Click Update tube

.

5. Use the bulb spin controls

to select the other bulb(s) for which

you noted calibration constant errors. Input the correct values for
each. Make certain to click Update Tube after you have corrected C
and E calibration values for each bulb BEFORE selecting the next
bulb.

6. When you have entered corrected constant values for each bulb on

both tubes, click

to exit the Calibration window.

You have verified the software configuration of VISCPRO

®

. To test

samples with your instrument, follow the instructions in Chapter 2. For
additional details regarding operating procedures for your instrument or
software, consult the appropriate section of this manual.

Checking Wash and Advanced Settings

To check Wash Settings for your miniQV-X instrument, click Configure
from the VISCPRO menu options and select the desired instrument. Then
click Tray Settings: Wash. The Tray Settings: Wash window will
appear.